Evidence used
- No CISA KEV confirmation is currently recorded.
- EPSS is 0.17% for the current model date.
BlackTreeCVE IntelligenceDräger · Zeus IE
High technical severity; prioritise exposed affected systems while verifying vendor guidance.
High technical severity; prioritise exposed affected systems while verifying vendor guidance.
Fix not verifiedDräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
The product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource.
An attacker operating through physical access may attempt exploitation without authentication or user interaction. If successful, the issue may cause the confidentiality, integrity or availability impact described by the vendor.
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
The product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource.
An attacker operating through physical access may attempt exploitation without authentication or user interaction. If successful, the issue may cause the confidentiality, integrity or availability impact described by the vendor.
CVSS severity, EPSS forecast probability, public exploit material and CISA-confirmed exploitation are separate signals.
No CISA KEV match was present at the last successful refresh. This means no confirmation from that source, not proof of no exploitation.
No exploit-tagged reference or CISA SSVC proof-of-concept state is currently recorded. Research may still exist outside the structured feeds.
CWE-668: Exposure of Resource to Wrong Sphere. The product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource.
CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:NCommon Vulnerability Scoring System 4.0: the compact vector below is decoded into plain language.
Operational remediation based on structured source evidence.
Published 2 Jun 2026 · Last source change 3 Jun 2026, 20:01 UTC · CWE-668 · Exposure of Resource to Wrong Sphere
Core structured fields are present and their contributing authorities are shown above.
No material field changes have been recorded since change tracking began. Routine source refreshes and cosmetic edits are intentionally excluded.