BlackTreeCVE IntelligenceMedtronic · InsterStim Applications
Medium technical severity with no CISA KEV confirmation; remediate through the normal risk-based patch cycle unless local exposure raises the priority.
Medium technical severity with no CISA KEV confirmation; remediate through the normal risk-based patch cycle unless local exposure raises the priority.
Patch availableMedtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
When setting a new password for a user, the product does not require knowledge of the original password, or using another form of authentication.
An attacker operating through physical access may attempt exploitation without authentication or user interaction. If successful, the issue may cause the confidentiality, integrity or availability impact described by the vendor.
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
When setting a new password for a user, the product does not require knowledge of the original password, or using another form of authentication.
An attacker operating through physical access may attempt exploitation without authentication or user interaction. If successful, the issue may cause the confidentiality, integrity or availability impact described by the vendor.
CVSS severity, EPSS forecast probability, public exploit material and CISA-confirmed exploitation are separate signals.
No CISA KEV match was present at the last successful refresh. This means no confirmation from that source, not proof of no exploitation.
No exploit-tagged reference or CISA SSVC proof-of-concept state is currently recorded. Research may still exist outside the structured feeds.
CWE-620: Unverified Password Change. When setting a new password for a user, the product does not require knowledge of the original password, or using another form of authentication.
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:HCommon Vulnerability Scoring System 3.1: the compact vector below is decoded into plain language.
Operational remediation based on structured source evidence.
Published 1 Mar 2023 · Last source change 7 Mar 2025, 21:34 UTC · CWE-620 · Unverified Password Change