Evidence used
- No CISA KEV confirmation is currently recorded.
- EPSS is 0.12% for the current model date.
BlackTreeCVE IntelligenceAliveCor · KardiaMobile
Medium technical severity with no CISA KEV confirmation; remediate through the normal risk-based patch cycle unless local exposure raises the priority.
Medium technical severity with no CISA KEV confirmation; remediate through the normal risk-based patch cycle unless local exposure raises the priority.
Fix not verifiedThe physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.
The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.
The product does not encrypt sensitive or critical information before storage or transmission.
An attacker operating through physical access may attempt exploitation without authentication or user interaction. If successful, the issue may cause the confidentiality, integrity or availability impact described by the vendor.
The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.
The product does not encrypt sensitive or critical information before storage or transmission.
An attacker operating through physical access may attempt exploitation without authentication or user interaction. If successful, the issue may cause the confidentiality, integrity or availability impact described by the vendor.
CVSS severity, EPSS forecast probability, public exploit material and CISA-confirmed exploitation are separate signals.
No CISA KEV match was present at the last successful refresh. This means no confirmation from that source, not proof of no exploitation.
No exploit-tagged reference or CISA SSVC proof-of-concept state is currently recorded. Research may still exist outside the structured feeds.
CWE-311: Missing Encryption of Sensitive Data. The product does not encrypt sensitive or critical information before storage or transmission.
CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:L/A:HCommon Vulnerability Scoring System 3.1: the compact vector below is decoded into plain language.
Operational remediation based on structured source evidence.
Published 27 Oct 2022 · Last source change 16 Apr 2025, 16:08 UTC · CWE-311 · Missing Encryption of Sensitive Data
Core structured fields are present and their contributing authorities are shown above.
No material field changes have been recorded since change tracking began. Routine source refreshes and cosmetic edits are intentionally excluded.