Evidence used
- No CISA KEV confirmation is currently recorded.
- The selected CVSS metric records a network-reachable, unauthenticated path with no user interaction.
- EPSS is 2.57% for the current model date.
BlackTreeCVE IntelligenceBaxter · SIGMA Spectrum Infusion System
Critical technical impact with a remotely reachable, unauthenticated path; no CISA KEV confirmation is currently recorded.
Critical technical impact with a remotely reachable, unauthenticated path; no CISA KEV confirmation is currently recorded.
Mitigation availableBaxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
This weakness has been deprecated because it covered redundant concepts already described in CWE-287.
An attacker operating through a network path may attempt exploitation without authentication or user interaction. If successful, the issue may cause the confidentiality, integrity or availability impact described by the vendor.
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
This weakness has been deprecated because it covered redundant concepts already described in CWE-287.
An attacker operating through a network path may attempt exploitation without authentication or user interaction. If successful, the issue may cause the confidentiality, integrity or availability impact described by the vendor.
CVSS severity, EPSS forecast probability, public exploit material and CISA-confirmed exploitation are separate signals.
No CISA KEV match was present at the last successful refresh. This means no confirmation from that source, not proof of no exploitation.
No exploit-tagged reference or CISA SSVC proof-of-concept state is currently recorded. Research may still exist outside the structured feeds.
CWE-592: DEPRECATED: Authentication Bypass Issues. This weakness has been deprecated because it covered redundant concepts already described in CWE-287.
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HCommon Vulnerability Scoring System 3.0: the compact vector below is decoded into plain language.
Operational remediation based on structured source evidence.
Published 26 Mar 2019 · Last source change 6 Aug 2024, 11:41 UTC · CWE-592 · DEPRECATED: Authentication Bypass Issues
Core structured fields are present and their contributing authorities are shown above.